Service Guide N5242-90001
Tests and Adjustments
PNA Series Microwave Network Analyzers
About System Verification and Performance Tests
N5241A/42A
About System Verification and Performance Tests
The performance of the network analyzer is specified in two ways: system specifications,
and instrument specifications. It is the end user’s responsibility to determine which set of
specifications is applicable to their use of the PNA.
A network analyzer measurement “system” includes the analyzer, calibration kit, test
cables, and any necessary adapters.
The system verification software in the PNA is
used to verify the system’s conformance to the “system” specifications.
A “pass”
result demonstrates that the analyzer, test cables, and adapters, perform correctly as a
system. It DOES NOT demonstrate that any one component performs according to its
individual specifications. A change to any part of this measurement system requires a
re-verification of the system.
Instrument specifications specify the network analyzer’s uncorrected measurement port
characteristics and its output and input behavior.
The PNA performance tests are
used to verify the analyzer’s conformance to “instrument” specifications.
System Specifications
System specifications specify warranted performance of the measurement system when
making error-corrected measurements using the same calibration kit and test cables used
during the system verification routine. System specifications are applicable only when the
measurement system is used to make error-corrected measurements.
The analyzer's system specifications are described in the
Agilent PNA Series Network
Analyzer Technical Specifications
and also in the analyzer’s on-line help system in the
section titled “Corrected System Performance”.
System specifications are expressed in two ways:
• residual errors of the measurement system shown as tabular specification values
• graphs of measurement uncertainty versus reflection and transmission coefficients
System specifications are verified in one of the following ways:
• Complete the system verification procedure using a certified verification kit and
certified calibration kit that will be used for future measurements, or
• Complete all of the performance tests using a certified calibration kit that will be used
for future measurements. This alternative verifies both the system specifications and
the instrument specifications for the analyzer.
Instrument Specifications
The analyzer's instrument specifications are described in the
Agilent PNA Series Network
Analyzer Technical Specifications
and also in the analyzer’s on-line help system in the
sections titled “Uncorrected System Performance”, “Test Port Output”, and “Test Port
Input”.
These specifications apply when the analyzer is used to make either raw or error-corrected
measurements.
Summary of Contents for N5241A
Page 11: ...Service Guide N5242 90001 1 1 1 Safety and Regulatory Information...
Page 19: ...Service Guide N5242 90001 2 1 2 General Product Information...
Page 35: ...Service Guide N5242 90001 3 1 3 Tests and Adjustments...
Page 81: ...Service Guide N5242 90001 4 1 4 Troubleshooting...
Page 125: ...Service Guide N5242 90001 5 1 5 Theory of Operation...
Page 159: ...Service Guide N5242 90001 6 1 6 Replaceable Parts...
Page 243: ...Service Guide N5242 90001 7 1 7 Repair and Replacement Procedures...
Page 313: ...Service Guide N5242 90001 8 1 8 General Purpose Maintenance Procedures...