Service Guide N5242-90001
4-29
PNA Series Microwave Network Analyzers
Troubleshooting
N5241A/42A
Measurement System Troubleshooting
Checking the Source Group
Source Group Tests
Before checking the source group assemblies, you must open the analyzer.
CAUTION
Use an antistatic work surface and wrist strap to reduce the chance of
electrostatic discharge for all of the procedures in this chapter.
1. Turn off the analyzer power.
2. Unplug the power to the analyzer and disconnect all front and rear panel connections
except installed jumpers.
3. Remove the outer and inner covers from the analyzer. Refer to
“Removing the Covers” on page 7-6
WARNING
Procedures described in this document are performed with power
supplied to the product while protective covers are removed. Energy
available at many points may, if contacted, result in personal injury.
4. With the covers off, plug in the analyzer and turn on the power.
CAUTION
Do not operate the analyzer with the outer cover removed for more than 30 minutes,
as this could cause the analyzer to overheat which could result in costly damage.
Frequency Banded vs. Broadband Failure
There are two main types of failures that are related to the source group. Frequency
banded failures are indicated by all receiver traces having partial dropouts across the
frequency range. Broadband failures are indicated by all receiver traces being in the noise
floor.
RF Signal Troubleshooting
Check the output power of the A, B, C, and D signals:
Equipment Used for This Check
Equipment Type
Model or
Part Number
Alternate Model or
Part Number
Power meter
E4418B/E4419B
E4418A/E4419A
Power sensor, 2.4 mm
E8487A
None
Adapter, 2.4 mm (f) to 2.4 mm (f)
11900B
85056-60007
Summary of Contents for N5241A
Page 11: ...Service Guide N5242 90001 1 1 1 Safety and Regulatory Information...
Page 19: ...Service Guide N5242 90001 2 1 2 General Product Information...
Page 35: ...Service Guide N5242 90001 3 1 3 Tests and Adjustments...
Page 81: ...Service Guide N5242 90001 4 1 4 Troubleshooting...
Page 125: ...Service Guide N5242 90001 5 1 5 Theory of Operation...
Page 159: ...Service Guide N5242 90001 6 1 6 Replaceable Parts...
Page 243: ...Service Guide N5242 90001 7 1 7 Repair and Replacement Procedures...
Page 313: ...Service Guide N5242 90001 8 1 8 General Purpose Maintenance Procedures...