Service Guide N5230-90014
8-17
PNA Series Microwave Network Analyzers
General Purpose Maintenance Procedures
N5230A
Option Enable Utility
Installing or Changing a Serial Number
It is necessary to reinstall the instrument serial number if the test set motherboard is
replaced.
IMPORTANT
Use extreme care when entering the serial number, as only one attempt is
allowed.
1. To change a serial number, select
Change
Serial
from the
Option
Enable
menu bar (see
). The current serial number is displayed. If no serial number
has previously been entered, the word “NONE” will be displayed.
2. VERY CAREFULLY, type the new serial number into the space provided and then click
Change
.
3. If an error is made in entering the serial number, obtain a clear code from Agilent, enter
the clear code in the space provided, and then click
Clear
. The correct serial number can
then be entered.
NOTE
To change an incorrect serial number, a clear-code password is required.
Contact Agilent to obtain the clear-code. Refer to
Summary of Contents for N5230A
Page 10: ...Contents 6 Service Guide N5230 90014 Contents...
Page 11: ...Service Guide N5230 90014 1 1 1 Safety and Regulatory Information...
Page 19: ...Service Guide N5230 90014 2 1 2 General Product Information...
Page 31: ...Service Guide N5230 90014 3 1 3 Tests and Adjustments...
Page 71: ...Service Guide N5230 90014 4 1 4 Troubleshooting...
Page 121: ...Service Guide N5230 90014 5 1 5 Theory of Operation...
Page 147: ...Service Guide N5230 90014 6 1 6 Replaceable Parts...
Page 183: ...Service Guide N5230 90014 7 1 7 Repair and Replacement Procedures...
Page 229: ...Service Guide N5230 90014 8 1 8 General Purpose Maintenance Procedures...