Service Guide N5230-90014
8-5
PNA Series Microwave Network Analyzers
General Purpose Maintenance Procedures
N5230A
Error Terms
Figure 8-2 Flowgraph of Two-Port Error Terms
where:
E = error term
1st Subscript:
D = Directivity
S = Source Match
R = Reflection Tracking
X = Crosstalk (Isolation)
L = Load Match
T = Transmission Tracking
2nd Subscript:
F = forward measurement (Port 1 to Port 2)
R = reverse measurement (Port 2 to Port 1)
Summary of Contents for N5230A
Page 10: ...Contents 6 Service Guide N5230 90014 Contents...
Page 11: ...Service Guide N5230 90014 1 1 1 Safety and Regulatory Information...
Page 19: ...Service Guide N5230 90014 2 1 2 General Product Information...
Page 31: ...Service Guide N5230 90014 3 1 3 Tests and Adjustments...
Page 71: ...Service Guide N5230 90014 4 1 4 Troubleshooting...
Page 121: ...Service Guide N5230 90014 5 1 5 Theory of Operation...
Page 147: ...Service Guide N5230 90014 6 1 6 Replaceable Parts...
Page 183: ...Service Guide N5230 90014 7 1 7 Repair and Replacement Procedures...
Page 229: ...Service Guide N5230 90014 8 1 8 General Purpose Maintenance Procedures...