Agilent N5161A/62A/81A/82A/83A MXG Signal Generators Service Guide
Assembly Replacement
A2 Vector Module (N5162A with Option 012 and N5182A Option 1EM with Option 012)
3-30
A2 Vector Module (N5162A with Option 012 and N5182A Option 1EM with Option 012)
Tools Required
•
T-10 driver
•
5/16” open-ended wrench
•
needle-nose pliers
Removal Procedure
for this procedure.
1.
”Assembly Installation and Removal Overview” on page 4-2
, and complete the
”Assembly Removal and Installation Procedure”
.
2.
Disconnect the power cord.
3.
Remove the covers from the signal generator. Refer to
.
4.
Remove the rear panel from the signal generator. Refer to
”Rear Panel (N5162A/82A)” on page 3-24
NOTE
The rear panel must be removed to allow the pin connector on the A2 Vector Module to be disconnected from the A3 RF assembly
without damaging the connector pins.
5.
Place the signal generator flat with the A2 Vector Module facing up and the rear panel facing you.
6.
Using the 5/16” open-ended wrench, disconnect W5, W16, W17, and W7.
7.
Use the needle-nose pliers to disconnect the following cables,
•
W19 (I Input) from A3J15
•
W20 (Q Input) from A3J16.
8.
Using the T-10 driver, remove the five screws (1) that connect the A2 Vector Module to the chassis.
9.
Place your thumbs under the BNC connectors and push up on the A2 Vector Module to release it from the pin connector.
10. Remove the A2 Vector Module.
Replacement Procedure
1.
Reverse the order of the removal procedure.
2.
Torque all T-10 screws to 9 in–lbs.
3.
Torque all BNC connectors to 8 in–lbs.
4.
Perform the post-repair performance tests that pertain to this replacement procedure, see
Chapter 4, Post-Repair Procedures and
.