Agilent N5161A/62A/81A/82A/83A MXG Signal Generators Service Guide
Assembly Replacement
USB Board
3-12
USB Board
Tools Required
•
T-10 driver
Removal Procedure
for this procedure.
1.
Disconnect the power cord.
2.
Remove the covers from the signal generator. Refer to
.
3.
Remove the front-panel from the signal generator. Refer to
.
4.
Using the T-10 driver, remove the screw (1) securing the USB board to the front-panel.
5.
Remove USB board
Replacement Procedure
1.
Reverse the order of the removal procedure.
2.
Torque screw to 9 in–lbs.
3.
Perform the post-repair performance tests that pertain to this replacement procedure, see
Chapter 4, Post-Repair Procedures and
.