Getting Started
Clock Synthesizer 13.5 GHz User Guide
13
Before switching on this instrument, make sure the supply voltage is in the
specified range.
This instrument has auto ranging line voltage input. Be sure the supply
voltage is within the specified range.
In an ESD-safe environment, carefully remove the N4963A. Install on a flat
surface with unobstructed air flow to the back panel. Plug the AC power cord
into the power converter module and a wall socket, then plug the converter
module into the N4963A.
1.2.2
Important Notes
•
Use ESD protection at all times when using the instrument
•
Review min/max specifications before applying input signals
•
Use only SMA-connectors on the RF ports
•
Use only BNC-connectors on the 10MHz reference ports
•
Use dust jackets on unused back panel connectors
•
Situate the instrument away from heat sources, do not block the fan
1.2.3
Performance Recommendations
1.
When using differential-mode connections, ensure the cables are
phase balanced
2.
Differential connectors may be used single-ended if the unused output
is terminated in 50
Ω
3.
Use high quality cables and connector savers (or adaptors)
4.
Keep cable lengths short and minimize number of cable bends
5.
Use a 7 to 10 in-lbs torque wrench when attaching connectors
1.3
Connect the Hardware
•
Connect power cord to adaptor to N4963A; plug in the power cord.
•
Connect the
Jittered
OUT
to your instrument transmit clock input using the
filters provided. Note: Filters are provided for 6-11.5 GHz and 9-13 GHz
operation.
•
Connect the
Reference
OUT
to your instrument receive clock input using
the filters provided. Note: Filters are provided for 6-11.5 GHz and 9-13 GHz
operation.
This Manual:http://www.manuallib.com/agilent/n4963a-clock-synthesizer-user-guide.html