Performing Time Domain Measurements
Response Types
Chapter 4
82
Response Types
Agilent multiport test systems are able to display the response of a device as if it were
stimulated with either a step or an impulse waveform. The algorithm that is used is robust
enough that it can determine either type of response, even for band-limited devices.
Furthermore, it is not necessary for the device to be able to operate at DC. Both Time Domain
Transmission (TDT) and Time Domain Reflection (TDR) parameters can be displayed on the
horizontal axis either as a time or a distance. It is also possible to enter a velocity factor so
that the electrical length corresponds with the physical length of the DUT.
The time scale of transmission parameters shows the propagation delay through the device.
The time scale of the reflection parameters shows the characteristics of the DUT at a certain
time delay into the device. Therefore, the reflection parameter shows the one-way travel time
into the device, not the round-trip propagation delay that other instruments may show.
On the vertical axis, the impulse response is a reflection or transmission coefficient displayed
on either a linear or a logarithmic scale. This parameter can be displayed as an absolute
number, or relative to a minimum or maximum value of the response.
Likewise, the step response can be displayed on either a linear or a logarithmic scale.
Alternatively, a reflection parameter can be displayed as impedance versus time rather than
as a reflection coefficient. Any of the single-ended or mixed-mode S-parameters can be
displayed in the time-domain. The table below shows the relationship between frequency
domain parameters to their time domain equivalents.
Relationship of Frequency Domain Parameters to Time Domain Equivalents
Mode
Direction
Type
Parameter
UNBALANCED
N/A
TDR
S11
UNBALANCED
REVERSE
TDT
S12
UNBALANCED
REVERSE
TDT
S13
UNBALANCED
REVERSE
TDT
S14
UNBALANCED
FORWARD
TDT
S21
UNBALANCED
N/A
TDR
S22
UNBALANCED
REVERSE
TDT
S23
UNBALANCED
REVERSE
TDT
S24
Summary of Contents for N4413A
Page 4: ...iv ...
Page 11: ...1 1 About the Multiport Test System ...
Page 18: ...About the Multiport Test System Available Options and Accessories Chapter 1 8 ...
Page 19: ...9 2 User Interface ...
Page 27: ...17 3 Using the Multiport Test System ...
Page 89: ...79 4 Performing Time Domain Measurements ...
Page 109: ...99 5 Menu Reference ...
Page 163: ...153 6 Maintenance and Troubleshooting ...
Page 172: ...Maintenance and Troubleshooting Contacting Agilent Chapter 6 162 ...
Page 173: ...163 7 Instrument Information ...
Page 208: ...Instrument Information Specifications and Characteristics Chapter 7 198 ...
Page 209: ...199 8 Safety and Regulatory Information ...
Page 218: ...Safety and Regulatory Information Regulatory Information Chapter 8 208 ...
Page 219: ...209 A Other Technical Procedures ...
Page 224: ...Other Technical Procedures IF Gain Adjustment Appendix A 214 ...