Using the Multiport Test System
How to Measure Simultaneous Conjugate Match of a Differential 2-Port Device
Chapter 3
71
How to Measure Simultaneous Conjugate Match of a
Differential 2-Port Device
The software includes a set of predefined display configurations, complete with the necessary
formulas, to help in measuring the simultaneous conjugate match of a differential two-port
device. To use these configurations:
1. Either load previously saved data (see
“File > Open > Data…” on page 101
), or make a new
measurement on a differential device.
2. Open the
Display Configuration
dialog box. See
“Setup > Display Configuration…” on
. Click
LOAD ALL
to open the simultaneous conjugate match display
configuration file, called Mixed Mode Parameters.mpc. This file is normally copied to
Multiport/User Config on installation.
3. Click
Set Display
for the Mixed Mode Parameters
Config Identifier
.
4. Click the
User-Defined
button on the main screen.
The Equations 1 and Equations 2 tabs on the user-defined display contain component
equations used in the Simultaneous Conj Match display.
Summary of Contents for N4413A
Page 4: ...iv ...
Page 11: ...1 1 About the Multiport Test System ...
Page 18: ...About the Multiport Test System Available Options and Accessories Chapter 1 8 ...
Page 19: ...9 2 User Interface ...
Page 27: ...17 3 Using the Multiport Test System ...
Page 89: ...79 4 Performing Time Domain Measurements ...
Page 109: ...99 5 Menu Reference ...
Page 163: ...153 6 Maintenance and Troubleshooting ...
Page 172: ...Maintenance and Troubleshooting Contacting Agilent Chapter 6 162 ...
Page 173: ...163 7 Instrument Information ...
Page 208: ...Instrument Information Specifications and Characteristics Chapter 7 198 ...
Page 209: ...199 8 Safety and Regulatory Information ...
Page 218: ...Safety and Regulatory Information Regulatory Information Chapter 8 208 ...
Page 219: ...209 A Other Technical Procedures ...
Page 224: ...Other Technical Procedures IF Gain Adjustment Appendix A 214 ...