Other Technical Procedures
Waveguide Characterization Procedure
Appendix A
211
Waveguide Characterization Procedure
Use this procedure to characterize a waveguide measurement.
1. Calibrate the Agilent microwave network analyzer.
2. Measure the DUT.
3. Save the measured data in S2P format on the network analyzer by pressing the
Save/Recall
key and then pressing the
DEFINE DISK-SAVE
,
SAVE USING ASCII
,
and
FORMAT ARY ON
off softkeys.
4. Import this measured data into the balanced measurement software by clicking
Import
and
Import S2P
on the
File
menu.
5. Export the data as a citifile by selecting
Export…
and
Citifile
on the
File
menu.
6. Open the Citifile in a text editor. The first three lines of the file are shown below.
CITIFILE A.01.00
NAME S-Parameters
VAR FREQ MAG 400
.
.
.
7. Insert the following two lines between the first and second lines of the Citifile.
#TXS TYPE=THRU
#TXS PORT1=WR-51 (W/G) PORT2=WR-51 (W/G)
8. Save this file and use it in the adapter section of the calibration.
Summary of Contents for N4413A
Page 4: ...iv ...
Page 11: ...1 1 About the Multiport Test System ...
Page 18: ...About the Multiport Test System Available Options and Accessories Chapter 1 8 ...
Page 19: ...9 2 User Interface ...
Page 27: ...17 3 Using the Multiport Test System ...
Page 89: ...79 4 Performing Time Domain Measurements ...
Page 109: ...99 5 Menu Reference ...
Page 163: ...153 6 Maintenance and Troubleshooting ...
Page 172: ...Maintenance and Troubleshooting Contacting Agilent Chapter 6 162 ...
Page 173: ...163 7 Instrument Information ...
Page 208: ...Instrument Information Specifications and Characteristics Chapter 7 198 ...
Page 209: ...199 8 Safety and Regulatory Information ...
Page 218: ...Safety and Regulatory Information Regulatory Information Chapter 8 208 ...
Page 219: ...209 A Other Technical Procedures ...
Page 224: ...Other Technical Procedures IF Gain Adjustment Appendix A 214 ...