3HUIRUPDQFH
7HVW&DUGLDF
2XWSXW
6WHS
Connect the patient simulator to the C.O.
Module using the patient cable.
6WHS
Configure the Patient simulator as
follows:
Injection temperature: 2 °C
Computation Const: 0.543
(Edward's Catheter)
Flow 5 l/min
6WHS
Check displayed value against the
simulator configuration.
C.O. = 5 +/– 1 l/min.
7KHVHUHVXOWV
GRQRWKDYHWR
EHUHSRUWHG
3HUIRUPDQFH
7HVW&DUGLDF
2XWSXW
FRQWG
6HUYLFH7RRO
SURFHGXUH
1RWH
This procedure applies for both Service Tools
M1012-14232C and M1012-61601.
6WHS
Make sure to be in Monitoring Mode.
Connect C.O. Interface Cable to module.
6WHS
Connect one side of the Service Tool to
Injectate receptacle of C.O. Interface
Cable and the other side to Catheter
Cable receptacle.
6WHS
Enter Measure C.O. Task Window and
check results.
Tblood = 37.0
o
C +/- 0.1
o
C
Tinj
=
0.0
o
C +/- 0.1
o
C
7KHVHUHVXOWV
GRQRWKDYHWR
EHUHSRUWHG
3HUIRUPDQFH
7HVW,QYDVLYH
3UHVVXUH
6WHS
Connect the patient simulator to the
Pressure module.
6WHS
Set patient simulator to 0 pressure.
6WHS
Make a Zero Calibration with the module.
6WHS
Configure the patient simulator for
P(static) = 200 mmHg.
6WHS
Wait for the display.
6WHS
Check displayed value against the
simulator configuration. If there is a
difference, calibrate the module.
Note:
If the Module was calibrated with a dedi-
cated reusable catheter, check the calibration
together with this catheter.
P = 200 +/– 5 mmHg
7KHVHUHVXOWV
GRQRWKDYHWR
EHUHSRUWHG
3HUIRUPDQFH
7HVW
7HPSHUDWXUH
6WHS
Connect the patient simulator to the
temperature input.
6WHS
Configure the patient simulator to 40 °C
or alternatively 100 °F.
6WHS
Check displayed value against the
simulator configuration.
+/– 0.2 °C
or +/– 0.4 °F
7KHVHUHVXOWV
GRQRWKDYHWR
EHUHSRUWHG
3HUIRUPDQFH
7HVW6S2
6WHS
Connect the sensor to the SpO
2
module.
6WHS
Measure the SpO
2
value on your own
finger (this assumes that you are healthy).
6WHS
Check displayed value against the
simulator configuration.
Between 95 and 100%
7KHVHUHVXOWV
GRQRWKDYHWR
EHUHSRUWHG
7DEOH7HVWDQG,QVSHFWLRQ5HTXLUHPHQWV
7HVW%ORFN
1DPH
7HVWRU³,QVSHFWLRQ´WR3HUIRUP
([SHFWHG7HVW5HVXOWV
:KDWWR5HFRUG
RQ6HUYLFH
5HFRUG
Summary of Contents for M1205A V24
Page 1: ...3DUW 1R 0 3ULQWHG XQH 7KLUG GLWLRQ JLOHQW 0 9 DQG 9 6HUYLFH XLGH...
Page 4: ......
Page 12: ...12 Contents...
Page 16: ...JLOHQW 0 9 9 6HULHV 6HUYLFH XLGH 2YHUYLHZ...
Page 214: ...JLOHQW 0 9 9 6HULHV 6HUYLFH XLGH 5HSDLULQJ WKH JLOHQW 0RGHO 9 Repairing the Agilent Model V24...
Page 301: ...JLOHQW 0 9 9 6HULHV 6HUYLFH XLGH 5HSODFHPHQW 3DUWV Replacement Parts 17 4 3 7 12 5 6 26...
Page 358: ...QGH...