$JLOHQW0$996HULHV6HUYLFH*XLGH
)URQW+RXVLQJ$VVHPEO\9&9& 9&79&7
F
r
on
t H
o
u
s
in
g
A
ssem
b
ly
V
2
4C
/V
26
C
& V
2
4C
T/
V2
6CT
5HPRYLQJWKH)ODW3DQHO
Remove the Front Housing Assembly. See "Removing the Front Housing Assembly" on
page 9-2.
Remove the Flat Panel Display Assembly. See "Removing the Flat Panel Display Assem-
bly" on page 9-6.
Remove the Board Shield. See "Removing the Board Shield" on page 9-5.
Remove the Flat Panel cable (w/ferrite block) from the Flat Panel Adapter Board. (It is
not necessary to remove the 3-Board Assembly.)
Turn the Flat Panel Display Assembly over and lay on a clean flat surface (Display side
up).
)LQJHUSULQWVDQGGLUWEHWZHHQWKH)ODW3DQHO'LVSOD\DQGSODVWLFDQWLJODUHVFUHHQVKRZ
XSSURPLQHQWO\GXULQJLQVWUXPHQWRSHUDWLRQ7DNHFDUHQRWWRWRXFKWKH)ODW3DQHO
VFUHHQRUWKHLQVLGHRIWKHFOHDUSODVWLFDQWLJODUHVFUHHQ5HPRYHDQ\ILQJHUSULQWVRUGLUW
XVLQJDQRQUHVLGXHIRUPLQJFOHDQHUVXFKDVDOFRKRODQGDOLQWIUHHFORWK
Remove 4 screws securing the Flat Panel to the Flat Panel Bracket
If the Flat Panel is being replaced, remove the ribbon cable with ferrite block from the Flat
Panel and move it to the replacement Flat Panel. The (4) connectors to the Flat Panel are
keyed and can be carefully removed with a small flat-bladed screwdriver.
&$87,21
Flat Panel Bracket
Flat Panel
127(
Summary of Contents for M1205A V24
Page 1: ...3DUW 1R 0 3ULQWHG XQH 7KLUG GLWLRQ JLOHQW 0 9 DQG 9 6HUYLFH XLGH...
Page 4: ......
Page 12: ...12 Contents...
Page 16: ...JLOHQW 0 9 9 6HULHV 6HUYLFH XLGH 2YHUYLHZ...
Page 214: ...JLOHQW 0 9 9 6HULHV 6HUYLFH XLGH 5HSDLULQJ WKH JLOHQW 0RGHO 9 Repairing the Agilent Model V24...
Page 301: ...JLOHQW 0 9 9 6HULHV 6HUYLFH XLGH 5HSODFHPHQW 3DUWV Replacement Parts 17 4 3 7 12 5 6 26...
Page 358: ...QGH...