To Remove a Column Module Assembly from the
LTM Oven Door
1
Gather the following:
•
Gloves, heat-resistant (for handling hot parts)
•
Gloves, lint free (to prevent contamination of the column, ferrules, and so forth with skin
oil and dirt)
•
Two 1/4-inch open-end wrenches
•
One 7/16-inch open end wrench
2
On the GC touchscreen, go to
Maintenance > Instrument > Perform Maintenance >
Maintenance Mode > Perform Maintenance
to cool down the GC ovens, inlets and
detectors.
3
.
4
Open the LTM oven door.
5
Using two wrenches, disconnect the In and Out segments from the CFT unions.
6
Loosen the retained screws that secure the column module assembly into the LTM oven
door.
Figure 35.
Two column modules installed in LTM oven door (shown after installation of unions)
7
Carefully slide the column assembly from the LTM oven door.
8
Unplug the LTM power cord.
9
Open the electronics enclosure by gently turning the small knob counterclockwise as
described in
“Connect the power supply and communications cables”
.
10
Disconnect the transfer line module and column module wires from the electronics board.