User-Defined Sequences
User-Defined Sequences - Concepts
This section describes the basics of user-defined sequences.
A sequence is created and maintained by means of the
Sequence Editor
that can
be enabled from the Pattern window.
A sequence consists of up to 120 blocks than can be looped. Each block can
generate a pause signal (constant 0 or 1), a divided clock signal, a 2^n -1 PRBS, or
a user pattern.
Single or multiple blocks can be looped. The sum of the blocks and the counted
loops must not exceed 120. An overall loop restarts the sequence after it has come
to its end.
When to Use a Sequence
You may wish to test a device that uses a certain protocol for processing data.
For example, the device might expect synchronization data, a preamble, payload
data, and a suffix.
All this can be provided by a user-defined sequence.
How a Sequence is Defined
The sequence is defined by a SequenceExpression which is formulated in its own
language, checked by the
Sequence Editor
.
The SequenceExpression specifies:
• the sequence start (and break) conditions
• the blocks, their contents, and trigger output
• the loops
You can inspect the contents of the SequenceExpression in the
Properties
dialog
of the Sequence Editor.
The SequenceExpression uses the following keywords:
Version=
optional
3
Setting up Patterns
96
Agilent J-BERT N4903B High-Performance Serial BERT
Summary of Contents for J-BERT N4903B
Page 1: ...S Agilent J BERT N4903B High Performance Serial BERT User Guide s Agilent Technologies ...
Page 10: ...10 Agilent J BERT N4903B High Performance Serial BERT ...
Page 36: ...1 Planning the Test 36 Agilent J BERT N4903B High Performance Serial BERT ...
Page 60: ...2 Setting up External Instrument s 60 Agilent J BERT N4903B High Performance Serial BERT ...
Page 120: ...3 Setting up Patterns 120 Agilent J BERT N4903B High Performance Serial BERT ...
Page 360: ...6 Advanced Analysis 360 Agilent J BERT N4903B High Performance Serial BERT ...
Page 468: ...8 Jitter Tolerance Tests 468 Agilent J BERT N4903B High Performance Serial BERT ...
Page 524: ...9 Solving Problems 524 Agilent J BERT N4903B High Performance Serial BERT ...
Page 566: ...10 Customizing the Instrument 566 Agilent J BERT N4903B High Performance Serial BERT ...