Table 10
n
Sequence Length
Longest Run of 1's Longest Run of 0's
7
128
7
7
10
1,024
10
10
11
2,048
11
11
13
8,192
13
13
15
32,768
15
15
23
8,388,608
23
23
Zero Substitution Patterns
A potential risk to bit errors are longer rows of zeros within a data stream. The
longest run of zeros in a 2^n PRBS consists of n zeros.
The Zero Substitution function can be used to stress the DUT additionally by
inserting a longer row of zeros in the data stream. The length of the zero row can
be freely specified, up to the length of the pattern.
The following figure shows an example, where a run of 10 zeroes is inserted into
a PRBS 2^7 pattern.
*
ZRUN
Run of 0s
000000
0
10
1
000100
000000100000
0
110
PRBN
ZSUB
* Closing bit
Mark Density Patterns
The mark density of a pattern is the portion of bits that are ones.
A pure PRBS pattern contains an equal number of ones and zeros and, thus, has
an average voltage in the center between high and low level. Therefore, the mark
density of such a pattern is 1/2.
Some devices may have problems processing data streams with different mark
density as this results in other average voltages and, thus, different electrical
conditions.
3
Setting up Patterns
92
Agilent J-BERT N4903B High-Performance Serial BERT
Summary of Contents for J-BERT N4903B
Page 1: ...S Agilent J BERT N4903B High Performance Serial BERT User Guide s Agilent Technologies ...
Page 10: ...10 Agilent J BERT N4903B High Performance Serial BERT ...
Page 36: ...1 Planning the Test 36 Agilent J BERT N4903B High Performance Serial BERT ...
Page 60: ...2 Setting up External Instrument s 60 Agilent J BERT N4903B High Performance Serial BERT ...
Page 120: ...3 Setting up Patterns 120 Agilent J BERT N4903B High Performance Serial BERT ...
Page 360: ...6 Advanced Analysis 360 Agilent J BERT N4903B High Performance Serial BERT ...
Page 468: ...8 Jitter Tolerance Tests 468 Agilent J BERT N4903B High Performance Serial BERT ...
Page 524: ...9 Solving Problems 524 Agilent J BERT N4903B High Performance Serial BERT ...
Page 566: ...10 Customizing the Instrument 566 Agilent J BERT N4903B High Performance Serial BERT ...