
In
Manual
mode, you can edit the automatically calculated frequencies. You can
also enter your own list of frequencies, with a minimum of two frequencies
specified. For example, you can jump between low and high jitter frequencies.
Enter the frequencies in Hz, separated by commas, for example:
2e4, 1.75e6, 6.334e5, ...
Enabling this compliance curve helps in ensuring that the device under test
complies with a certain standard. This feature is available only if the compliance
jitter measurement option is installed. The standard is taken from the compliance
measurement.
BER Specification
The
Target Bit Error Ratio
defines the required BER. The jitter tolerance curve is
constructed from the points where the measured BER crosses the
Target Bit Error
Ratio
.
The
Verification Method
defines under which circumstances the measurement
proceeds from one jitter amplitude to the next. You can choose between two
Verification Methods
:
Setting a
Confidence Level
is generally recommended. This ensures that neither
too few nor too many bits are captured. See also
.
If you enable
Absolute Bits/Errors
, the measurement proceeds to the next jitter
amplitude as soon as either the number of received bits or the number of errors is
reached.
N O T E
Ensure that you capture a number of bits that suffices to verify (with the desired
probability) that a measured point has a BER below the
Target Bit Error Ratio.
This is the wait time that takes effect after the measured BER is higher than the
target BER. It allows the DUT to recover from the situation. The unit is seconds.
The next measurement starts after that time.
Search Specification
On this page of the
Jitter Tolerance Properties
dialog, you specify how the
measurement proceeds upwards or downwards from one measured point to the
8
Jitter Tolerance Tests
444
Agilent J-BERT N4903B High-Performance Serial BERT
Manual mode
Show Compliance Curve
Target Bit Error Ratio
Confidence Level
Absolute Bits/Errors
Relax Time
Summary of Contents for J-BERT N4903B
Page 1: ...S Agilent J BERT N4903B High Performance Serial BERT User Guide s Agilent Technologies ...
Page 10: ...10 Agilent J BERT N4903B High Performance Serial BERT ...
Page 36: ...1 Planning the Test 36 Agilent J BERT N4903B High Performance Serial BERT ...
Page 60: ...2 Setting up External Instrument s 60 Agilent J BERT N4903B High Performance Serial BERT ...
Page 120: ...3 Setting up Patterns 120 Agilent J BERT N4903B High Performance Serial BERT ...
Page 360: ...6 Advanced Analysis 360 Agilent J BERT N4903B High Performance Serial BERT ...
Page 468: ...8 Jitter Tolerance Tests 468 Agilent J BERT N4903B High Performance Serial BERT ...
Page 524: ...9 Solving Problems 524 Agilent J BERT N4903B High Performance Serial BERT ...
Page 566: ...10 Customizing the Instrument 566 Agilent J BERT N4903B High Performance Serial BERT ...