You can click on the corrosponding button to choose among the different modes
of Periodic Jitter 2. The green LED on the button indicates the currently selected
mode of Periodic Jitter 2.
The ability of sweeping the PJ settings supports testing DUTs according to the
PCIe 2.0 and PCIe 3.0 standard without the need of providing a jitter modulation
signal externally.
Periodic Jitter 2 – Constant Jitter Parameters
The Periodic Jitter 2 – Constant Jitter is characterized by:
• Amplitude
• Frequency
• Waveform
N O T E
You will not see the parameter (Amplitude and Frequency) in case the PJ Sweep is
enabled.
N O T E
The setting for the jitter Amplitude is shared with the Constant Amplitude Sweep
mode.
A graph indicates whether the chosen setting of frequency and amplitude is
tolerable.
8
Jitter Tolerance Tests
422
Agilent J-BERT N4903B High-Performance Serial BERT
Summary of Contents for J-BERT N4903B
Page 1: ...S Agilent J BERT N4903B High Performance Serial BERT User Guide s Agilent Technologies ...
Page 10: ...10 Agilent J BERT N4903B High Performance Serial BERT ...
Page 36: ...1 Planning the Test 36 Agilent J BERT N4903B High Performance Serial BERT ...
Page 60: ...2 Setting up External Instrument s 60 Agilent J BERT N4903B High Performance Serial BERT ...
Page 120: ...3 Setting up Patterns 120 Agilent J BERT N4903B High Performance Serial BERT ...
Page 360: ...6 Advanced Analysis 360 Agilent J BERT N4903B High Performance Serial BERT ...
Page 468: ...8 Jitter Tolerance Tests 468 Agilent J BERT N4903B High Performance Serial BERT ...
Page 524: ...9 Solving Problems 524 Agilent J BERT N4903B High Performance Serial BERT ...
Page 566: ...10 Customizing the Instrument 566 Agilent J BERT N4903B High Performance Serial BERT ...