These settings are used for data collection. Changes require to run the test
again. See
–
Pass/Fail
tab
These settings determine whether the calculated results are recognized as
passed or failed. However, a new test run is not required when doing
changes here. See
–
View
tab,
Graph
tab, and
Color
tab
All settings on these tabs only affect the way the data is displayed. You do
not need to run the measurement again. See
,
“How to Change the Colors of the Graph”
for details.
3
Press
OK
when you have made all required changes to close the
Properties
dialog box.
Spectral Jitter - Reference
The Spectral Jitter measurement returns the results in a graphical and in a
numerical form. The following sections provide explanations of the measured
parameters and the display options that are specific to this measurement.
Additionally, some information is provided to explain the theoretical background
behind the jitter calculations.
Properties that can be specified on the various tabs of the
Properties
dialog box:
Parameters Tab
N O T E
If you modify the parameters on this page, you have to rerun the measurement to
update the results.
Set the data acquisition parameters:
•
Acquisition Depth
Suitable numbers can be chosen from the drop-down list.
The
Acquisition Depth
determines the length of the time record used for the
FFT.
A smaller number reduces the duration of the Spectral Jitter measurement and
its precision. A larger number increases the measurement duration but also
the frequency resolution of the measured spectral components.
The frequency resolution of the measurement is:
Advanced Analysis
6
Agilent J-BERT N4903B High-Performance Serial BERT
353
Summary of Contents for J-BERT N4903B
Page 1: ...S Agilent J BERT N4903B High Performance Serial BERT User Guide s Agilent Technologies ...
Page 10: ...10 Agilent J BERT N4903B High Performance Serial BERT ...
Page 36: ...1 Planning the Test 36 Agilent J BERT N4903B High Performance Serial BERT ...
Page 60: ...2 Setting up External Instrument s 60 Agilent J BERT N4903B High Performance Serial BERT ...
Page 120: ...3 Setting up Patterns 120 Agilent J BERT N4903B High Performance Serial BERT ...
Page 360: ...6 Advanced Analysis 360 Agilent J BERT N4903B High Performance Serial BERT ...
Page 468: ...8 Jitter Tolerance Tests 468 Agilent J BERT N4903B High Performance Serial BERT ...
Page 524: ...9 Solving Problems 524 Agilent J BERT N4903B High Performance Serial BERT ...
Page 566: ...10 Customizing the Instrument 566 Agilent J BERT N4903B High Performance Serial BERT ...