Instead of UI (one
Unit Interval
is equal to one system clock period), the relative
time can also be specified in seconds.
• The
Voltage
is the decision threshold voltage at this measurement point. The
voltages of the measurement points can be set as absolute voltages, as offset
voltages, or as percentages. This is done on the
Parameters
page of the
Properties
dialog.
By default, the six measurement points are symmetrically placed, as illustrated in
the figure below:
These settings can be changed, and up to 32 measurement points can be defined.
N O T E
Critical areas at the error detector are generally close to ±0.5 UI.
Fast Eye Mask - Procedures
This section shows how to set up and perform a Fast Eye Mask measurement. As
an example we measure the eye of a shielded cable.
This requires the following steps:
• Preparing the Fast Eye Mask Measurement. (See
• Executing the Fast Eye Mask Measurement. (See
Eye Mask Measurement” on page 310
).
Advanced Analysis
6
Agilent J-BERT N4903B High-Performance Serial BERT
309
Summary of Contents for J-BERT N4903B
Page 1: ...S Agilent J BERT N4903B High Performance Serial BERT User Guide s Agilent Technologies ...
Page 10: ...10 Agilent J BERT N4903B High Performance Serial BERT ...
Page 36: ...1 Planning the Test 36 Agilent J BERT N4903B High Performance Serial BERT ...
Page 60: ...2 Setting up External Instrument s 60 Agilent J BERT N4903B High Performance Serial BERT ...
Page 120: ...3 Setting up Patterns 120 Agilent J BERT N4903B High Performance Serial BERT ...
Page 360: ...6 Advanced Analysis 360 Agilent J BERT N4903B High Performance Serial BERT ...
Page 468: ...8 Jitter Tolerance Tests 468 Agilent J BERT N4903B High Performance Serial BERT ...
Page 524: ...9 Solving Problems 524 Agilent J BERT N4903B High Performance Serial BERT ...
Page 566: ...10 Customizing the Instrument 566 Agilent J BERT N4903B High Performance Serial BERT ...