This graph shows two straight lines. The intersection of these lines marks the Q-
factor and the Q optimum Threshold.
Mathematically, the standard deviation and mean values are calculated as:
s
= -
1
B
m
=
-A
B
This calculation leads to the values of μ
1,0
and σ
1,0
.
Notes on the Q-Results
You can specify the range of data points used for these calculations by means of
the
Min BER for Q
and
BER Threshold
parameters (see also
).
For example, you can calculate the Q-factor from relatively few data points with
measured bit error rates between 1e-10 and 1e-5. The calculation tells you the
optimum decision threshold and also the BER expected when the error detector
uses this threshold.
To measure such small bit error rates, you have to compare an adequate number
of bits. Depending on the DUT characteristics, you may also have to use a very fine
step size for the threshold variation.
On the other hand, such a comparison may make it easier to understand the
dependencies.
If you set the
BER Threshold
to a position that gives you only a few data points,
you will find:
• The Q-factor is relatively high.
Advanced Analysis
6
Agilent J-BERT N4903B High-Performance Serial BERT
285
Use of the Q-factor Calculations
Comparison of the Results
Summary of Contents for J-BERT N4903B
Page 1: ...S Agilent J BERT N4903B High Performance Serial BERT User Guide s Agilent Technologies ...
Page 10: ...10 Agilent J BERT N4903B High Performance Serial BERT ...
Page 36: ...1 Planning the Test 36 Agilent J BERT N4903B High Performance Serial BERT ...
Page 60: ...2 Setting up External Instrument s 60 Agilent J BERT N4903B High Performance Serial BERT ...
Page 120: ...3 Setting up Patterns 120 Agilent J BERT N4903B High Performance Serial BERT ...
Page 360: ...6 Advanced Analysis 360 Agilent J BERT N4903B High Performance Serial BERT ...
Page 468: ...8 Jitter Tolerance Tests 468 Agilent J BERT N4903B High Performance Serial BERT ...
Page 524: ...9 Solving Problems 524 Agilent J BERT N4903B High Performance Serial BERT ...
Page 566: ...10 Customizing the Instrument 566 Agilent J BERT N4903B High Performance Serial BERT ...