After this number of errors, the measurement stops for the current sample point
and moves to the next one. This allows you to speed up the measurement. You
can switch off this option if only the number of compared bits is important.
N O T E
The measurement moves to the next sample point when the first of the two criteria
is reached.
Set the criteria for the sample threshold:
•
Resolution
Specifies the distance between the sampling points. The lower this value is,
the more sampling points you have in the selected voltage range. The minimum
step width is hardware-dependent. At the time being, the minimum is 1 mV.
•
Low Level
This is the lower end of the measured voltage range.
•
High Level
This is the upper end of the measured voltage range.
•
Edge Resolution Optimization
Turns the resolution optimization on or off. If this option is enabled, the Serial
BERT intelligently sets a resolution so that there are more sampling points at
the edges. This can greatly improve the results without dramatically increasing
the duration of the test.
In the following example, a step width of 2 mV was used. This has led to 502
data points per Volt.
If
Edge Resolution Optimization
is enabled, the software first scans the
threshold range in relatively coarse steps and determines the areas where the
signal produces no edges-that means, where the BER is nearly constant.
Then, the algorithm uses the specified step width for measuring the regions
where the BER changes, and a much lower resolution (about one tenth) for
measuring the regions where the BER is constant.
This reduces the number of data points to 145, as shown in the figure below.
Advanced Analysis
6
Agilent J-BERT N4903B High-Performance Serial BERT
271
Summary of Contents for J-BERT N4903B
Page 1: ...S Agilent J BERT N4903B High Performance Serial BERT User Guide s Agilent Technologies ...
Page 10: ...10 Agilent J BERT N4903B High Performance Serial BERT ...
Page 36: ...1 Planning the Test 36 Agilent J BERT N4903B High Performance Serial BERT ...
Page 60: ...2 Setting up External Instrument s 60 Agilent J BERT N4903B High Performance Serial BERT ...
Page 120: ...3 Setting up Patterns 120 Agilent J BERT N4903B High Performance Serial BERT ...
Page 360: ...6 Advanced Analysis 360 Agilent J BERT N4903B High Performance Serial BERT ...
Page 468: ...8 Jitter Tolerance Tests 468 Agilent J BERT N4903B High Performance Serial BERT ...
Page 524: ...9 Solving Problems 524 Agilent J BERT N4903B High Performance Serial BERT ...
Page 566: ...10 Customizing the Instrument 566 Agilent J BERT N4903B High Performance Serial BERT ...