Table 26
Parameter
Description
Definition
Optimal Sample Point
Delay
The average of the left (A)
and right (B) bathtub/BER
threshold intersections.
(A+B)/2
Phase Margin
The period of time where
the bit error rate is lower
than the BER threshold.
B-A
The A and B values are the left and right intersections of the bathtub curves with
the BER threshold. Obviously, all values change if the BER threshold is modified.
The following illustration shows an example for a measurement.
N O T E
This graphical view is not available in the measurement software.
The jitter measurement parameters are defined in the following list.
N O T E
All jitter measurement parameters (except the Total Jitter RMS and Total Jitter
Mean) change with the BER threshold.
•
Total Jitter RMS
The average of the left and right jitter histogram root mean squared values.
•
Total Jitter Peak-to-Peak
Peak-to-peak value for total jitter. Calculated as the pulse period (unit interval)
minus the
Phase Margin
.
6
Advanced Analysis
260
Agilent J-BERT N4903B High-Performance Serial BERT
Jitter Measurement Parameters
Summary of Contents for J-BERT N4903B
Page 1: ...S Agilent J BERT N4903B High Performance Serial BERT User Guide s Agilent Technologies ...
Page 10: ...10 Agilent J BERT N4903B High Performance Serial BERT ...
Page 36: ...1 Planning the Test 36 Agilent J BERT N4903B High Performance Serial BERT ...
Page 60: ...2 Setting up External Instrument s 60 Agilent J BERT N4903B High Performance Serial BERT ...
Page 120: ...3 Setting up Patterns 120 Agilent J BERT N4903B High Performance Serial BERT ...
Page 360: ...6 Advanced Analysis 360 Agilent J BERT N4903B High Performance Serial BERT ...
Page 468: ...8 Jitter Tolerance Tests 468 Agilent J BERT N4903B High Performance Serial BERT ...
Page 524: ...9 Solving Problems 524 Agilent J BERT N4903B High Performance Serial BERT ...
Page 566: ...10 Customizing the Instrument 566 Agilent J BERT N4903B High Performance Serial BERT ...