Understanding Bit Recovery Mode
In bit recovery mode (BRM), the error detector does not expect any particular
pattern. Nevertheless, it detects bit errors.
In bit recovery mode, the error detector uses two sampling points. The second
sampling point is not visible. Both are set to the expected optimum position when
the
Auto Align
button is pressed (with respect to clock rate, signal voltages and
offset, decision threshold, and so on).
The bits received at this sampling point are considered to be correct.
You can now move the visible sampling point and investigate the behavior of the
DUT in detail. The error detector compares the bits at your sampling point with the
bits recognized at the hidden sampling point. As you move the visible sampling
point, the indicated BER will increase.
N O T E
The indicated BER is a relative BER. It approaches the absolute BER if the error ratio
at the initial sampling point is very low. It is identical with the true BER if the signal
at the initial sampling point is error free.
In bit recovery mode, some restrictions apply:
• Synchronization is impossible, because no particular data is expected.
Sync
now
and
auto synchronization
functions are disabled.
• Burst mode is disabled.
Setting up the Error Detector
5
Agilent J-BERT N4903B High-Performance Serial BERT
195
Summary of Contents for J-BERT N4903B
Page 1: ...S Agilent J BERT N4903B High Performance Serial BERT User Guide s Agilent Technologies ...
Page 10: ...10 Agilent J BERT N4903B High Performance Serial BERT ...
Page 36: ...1 Planning the Test 36 Agilent J BERT N4903B High Performance Serial BERT ...
Page 60: ...2 Setting up External Instrument s 60 Agilent J BERT N4903B High Performance Serial BERT ...
Page 120: ...3 Setting up Patterns 120 Agilent J BERT N4903B High Performance Serial BERT ...
Page 360: ...6 Advanced Analysis 360 Agilent J BERT N4903B High Performance Serial BERT ...
Page 468: ...8 Jitter Tolerance Tests 468 Agilent J BERT N4903B High Performance Serial BERT ...
Page 524: ...9 Solving Problems 524 Agilent J BERT N4903B High Performance Serial BERT ...
Page 566: ...10 Customizing the Instrument 566 Agilent J BERT N4903B High Performance Serial BERT ...