3
In the
Aux In
section, define whether the Serial BERT should insert pattern B
in
Level Sensitive
or
Edge Sensitive
mode.
for more information on the different available
modes.
4
Click
OK
to close the dialog box.
Suppressing the Outputs via External Signal
You can configure the Serial BERT to suppress the output according to an external
signal:
1
Connect an external instrument to the Aux In port.
The signals received at this port must be TTL-compatible. The granularity for
the insertion of pattern B via external signal is 512 bits.
2
Click
Alt. Pattern and Aux In
menu item from the
PG Setup
submenu to open
the
Alternate Pattern Control
dialog box.
3
In the
Aux In
section, select
Output Blanking
.
for more information on the different available
modes.
4
Click
OK
to close the dialog box.
Pattern Alternation - Reference
The Serial BERT offers the following options for alternating between pattern A and
pattern B:
N O T E
These functions are only available if
Alternate Pattern is selected in the Pattern
Properties dialog box.
Continuous A
Select this option to send out only pattern A repeatedly.
A0
A1
A2
A A A A A A A A A A A A
Data Out
4
Setting up the Pattern Generator
164
Agilent J-BERT N4903B High-Performance Serial BERT
Summary of Contents for J-BERT N4903B
Page 1: ...S Agilent J BERT N4903B High Performance Serial BERT User Guide s Agilent Technologies ...
Page 10: ...10 Agilent J BERT N4903B High Performance Serial BERT ...
Page 36: ...1 Planning the Test 36 Agilent J BERT N4903B High Performance Serial BERT ...
Page 60: ...2 Setting up External Instrument s 60 Agilent J BERT N4903B High Performance Serial BERT ...
Page 120: ...3 Setting up Patterns 120 Agilent J BERT N4903B High Performance Serial BERT ...
Page 360: ...6 Advanced Analysis 360 Agilent J BERT N4903B High Performance Serial BERT ...
Page 468: ...8 Jitter Tolerance Tests 468 Agilent J BERT N4903B High Performance Serial BERT ...
Page 524: ...9 Solving Problems 524 Agilent J BERT N4903B High Performance Serial BERT ...
Page 566: ...10 Customizing the Instrument 566 Agilent J BERT N4903B High Performance Serial BERT ...