having to care for the correct setup of the required data streams. The PG pattern
is considered to be the serialized output of the connected multiplexer.
De-Emphasis
In De-emphasis mode, an external connected passive resistor network adds two
bit streams. The bit streams provided at ‘Data Out’ and ‘Aux Out’ are identical
except for the bit stream at Aux Out which is shifted by one bit. The bitshift value
can only be -1 (Pre-Cursor) or +1 (Post-Cursor). For more information, see
“Controlling De-Emphasis Mode” on page 134
Delay Control Input
Delay Control Input - Concepts
The Serial BERT features a wide bandwidth voltage controlled delay line in its
pattern generator data path, which can be used to delay the data relative to the
clock. The delay is controlled by an external signal connected to the Delay Ctrl In
connector on the pattern generator front panel.
The delay control input is intended primarily for jitter tolerance testing, where a
data signal with a defined amount of jitter is needed. Random jitter (RJ) can be
created by applying high bandwidth white Gaussian noise, periodic deterministic
jitter (DJ) with sinusoidal (SJ) or rectangular/triangle signals (PJ).
The delay control input is most useful for the generation of high frequency, low
amplitude jitter. Higher amounts of jitter at lower frequencies (wander) can be
generated by supplying an FM modulated clock to the pattern generator's clock
input (Clk In) in external clock mode.
N O T E
With a Serial BERT on which the calibrated and integrated jitter injection option J10
is installed, the delay control input is rarely used as explained above. The option
J10 includes all the jitter sources required for jitter tolerance tests. For details see
“Jitter Tolerance Tests - Concepts ” on page 395
.
The delay control input is equipped with overvoltage protection. If the applied
voltage is too high, the input is automatically disabled, and the following dialog
appears:
Setting up the Pattern Generator
4
Agilent J-BERT N4903B High-Performance Serial BERT
155
Summary of Contents for J-BERT N4903B
Page 1: ...S Agilent J BERT N4903B High Performance Serial BERT User Guide s Agilent Technologies ...
Page 10: ...10 Agilent J BERT N4903B High Performance Serial BERT ...
Page 36: ...1 Planning the Test 36 Agilent J BERT N4903B High Performance Serial BERT ...
Page 60: ...2 Setting up External Instrument s 60 Agilent J BERT N4903B High Performance Serial BERT ...
Page 120: ...3 Setting up Patterns 120 Agilent J BERT N4903B High Performance Serial BERT ...
Page 360: ...6 Advanced Analysis 360 Agilent J BERT N4903B High Performance Serial BERT ...
Page 468: ...8 Jitter Tolerance Tests 468 Agilent J BERT N4903B High Performance Serial BERT ...
Page 524: ...9 Solving Problems 524 Agilent J BERT N4903B High Performance Serial BERT ...
Page 566: ...10 Customizing the Instrument 566 Agilent J BERT N4903B High Performance Serial BERT ...