Adjust Output Levels (optional)
Data, Aux Data, Clock and Trigger/Ref Clock offset and voltage levels can be
adjusted. This is typically done when you want to tune your BER measurement or
stress the device.
You can use the knobs on the Serial BERT's front panel to fine-tune the data and
clock amplitudes and offsets. If you want to set a specific value, you can use the
numeric keyboard.
To enter specific values for the outputs from the keyboard:
1
From the
PG Setup
menu, press
PG Data Setup
to modify Data and Aux Data,
and
PG Clock/Trig Setup
to modify Clock and Trigger.
2
Click into the field you want to change.
3
Enter the desired voltage using the numeric keyboard or the respective front
panel knob (only for Data and Clock).
N O T E
If the termination voltage is wrongly adjusted or the output ports are not connected,
an internal protection circuit automatically sets the output voltage levels to safe
levels. Connect the output ports or change the termination voltage before adjusting
the output levels.
“Understanding the Output Level Parameters” on page 126
for information
about how the Serial BERT works with output level parameters.
Controlling the De-Emphasis Signal Converter
After the
De-Emphasis
function has been enabled using the
External Instrument
(s) - Config
menu, the following parameters are visible in PG Data Output screen.
Controlling the N4916A De-Emphasis Signal Converter
For N4916A, you have de-emphasis amplitude (post–cursor) and the corresponding
unit.
4
Setting up the Pattern Generator
132
Agilent J-BERT N4903B High-Performance Serial BERT
Summary of Contents for J-BERT N4903B
Page 1: ...S Agilent J BERT N4903B High Performance Serial BERT User Guide s Agilent Technologies ...
Page 10: ...10 Agilent J BERT N4903B High Performance Serial BERT ...
Page 36: ...1 Planning the Test 36 Agilent J BERT N4903B High Performance Serial BERT ...
Page 60: ...2 Setting up External Instrument s 60 Agilent J BERT N4903B High Performance Serial BERT ...
Page 120: ...3 Setting up Patterns 120 Agilent J BERT N4903B High Performance Serial BERT ...
Page 360: ...6 Advanced Analysis 360 Agilent J BERT N4903B High Performance Serial BERT ...
Page 468: ...8 Jitter Tolerance Tests 468 Agilent J BERT N4903B High Performance Serial BERT ...
Page 524: ...9 Solving Problems 524 Agilent J BERT N4903B High Performance Serial BERT ...
Page 566: ...10 Customizing the Instrument 566 Agilent J BERT N4903B High Performance Serial BERT ...