a browser and search for the file that contains the pattern. You can search for
files with the extentions .ptrn, .txt, or .dat.
If the pattern file contains two alternate patterns, you can also specify whether
the A or B pattern shall be generated (see also
“Pattern File Specifications” on
). If the pattern file contains just one pattern, your choice is ignored.
You cannot change the block length. It shows the length of the pattern as stored
in the file.
If the pattern length is not 512 bits or a multiple of 512 bits and this single block
is looped, the behavior depends on the loop end condition.
If the loop end condition is INFinite or determined by the Aux In port, then the
pattern is automatically repeated until 512 bits or a multiple thereof is reached.
In this case, the duration of the block will be longer than indicated. For example,
a pattern containing 17 bits will be repeated 512 times and the actual block
length will be 8704 bits. This has an impact on the trigger at block start.
You can also use the
Select User Pattern
dialog to create new pattern.
If this checkbox is enabled, a trigger spike will be output at the pattern generator's
Trigger Out connector whenever the execution of the block begins, provided that
the Trigger Out port is set to
Sequence Trigger
mode (see also
).
Click this button to launch
Create New Pattern
dialog. See
Click this button to open the
Edit Pattern Window for Sequencer-Pattern Editor
.
“Edit Pattern Window for Sequencer-Pattern Editor” on page 110
Loops Within a Sequence
A loop defines the transition from the end of a block to the beginning of the same
or a previous block. It is not possible to jump into an existing loop. It is also not
possible to specify loops within loops (exept the default overall loop).
For information on how to create a loop, see
“Creating a New Sequence” on page
.
Setting up Patterns
3
Agilent J-BERT N4903B High-Performance Serial BERT
103
Block Trigger
Create Pattern
Edit Pattern(s)
Summary of Contents for J-BERT N4903B
Page 1: ...S Agilent J BERT N4903B High Performance Serial BERT User Guide s Agilent Technologies ...
Page 10: ...10 Agilent J BERT N4903B High Performance Serial BERT ...
Page 36: ...1 Planning the Test 36 Agilent J BERT N4903B High Performance Serial BERT ...
Page 60: ...2 Setting up External Instrument s 60 Agilent J BERT N4903B High Performance Serial BERT ...
Page 120: ...3 Setting up Patterns 120 Agilent J BERT N4903B High Performance Serial BERT ...
Page 360: ...6 Advanced Analysis 360 Agilent J BERT N4903B High Performance Serial BERT ...
Page 468: ...8 Jitter Tolerance Tests 468 Agilent J BERT N4903B High Performance Serial BERT ...
Page 524: ...9 Solving Problems 524 Agilent J BERT N4903B High Performance Serial BERT ...
Page 566: ...10 Customizing the Instrument 566 Agilent J BERT N4903B High Performance Serial BERT ...