4
Specify the data to be generated during execution of a block.
– Click the
Edit
button of the block.
– Choose from the list. For example, you can select a PRBS
polynomial or browse for a pattern file. A "None" block is
completely ignored.
– If necessary, change the block length. The block length of user
patterns is derived from the file. Remember that such patterns
are automatically repeated until a multiple of 512 bits is reached.
– Decide on Trigger On or Off. This setting becomes effective if the
Trigger Out port is set to
Sequence Trigger
(trigger at block begin).
5
Create loops (if desired):
– Click the out-arrow of a block. A loop condition box labeled INF
appears.
– Click the in-arrow of the same or a previous block. This closes the
loop.
– Click the condition box and choose the appropriate break
condition.
6
Download the sequence to the pattern generator and test it.
7
Save the sequence in a file for later re-use.
Saving the Sequence in a File
To save the present sequence under a new name:
1
In the toolbar, click the
Save As
icon.
2
Select a suitable directory, for example
C:\N4903a\Sequences
.
3
Enter a file name.
The filename extension is
.seq
by default.
If the present sequence was loaded from a file and edited, you can also
click the
Save
icon. In this case, the saved file is updated.
Loading a Sequence From a File
To recall a sequence from a file:
1
In the toolbar, click the
Open
icon.
2
Setting up Patterns
60
Agilent J-BERT N4903 High-Performance Serial BERT
Summary of Contents for J-BERT N4903
Page 1: ...S Agilent J BERT N4903 High Performance Serial BERT User Guide s Agilent Technologies...
Page 68: ...2 Setting up Patterns 68 Agilent J BERT N4903 High Performance Serial BERT...
Page 158: ...4 Setting up the Error Detector 158 Agilent J BERT N4903 High Performance Serial BERT...
Page 314: ...6 Evaluating Results 314 Agilent J BERT N4903 High Performance Serial BERT...
Page 374: ...7 Jitter Tolerance Tests 374 Agilent J BERT N4903 High Performance Serial BERT...
Page 394: ...8 Solving Problems 394 Agilent J BERT N4903 High Performance Serial BERT...
Page 434: ...Index 434 Agilent J BERT N4903 High Performance Serial BERT...