Consider these recommendations when composing the desired total
jitter (see
“Jitter Setup - Procedures” on page 326
sinusoidal jitter is automatically generated and no setup is required.
Frequency Specification
You can choose
Auto
or
Manual
mode.
In
Auto
mode, the jitter frequency is automatically swept from
Start
Frequency
to
Stop Frequency
.
•
Start Frequency
: This is the minimum and first jitter frequency to be
generated. It determines the left-hand border of the graphical result
display.
•
Stop Frequency
: This is the maximum and last jitter frequency to be
generated. It determines the right-hand border of the graphical
result display.
The measurement always includes the start and stop frequencies.
The frequency scale is logarithmic.
•
Number of Steps
: The
Number of Steps
between start and stop defines
the test frequencies.
The location of the individual test frequencies depends on the setting
of the
Auto Frequencies
checkbox on the
Standard
tab:
– Checkbox is selected (default):
The list yields frequencies that are recommended for the selected
standard, for example, the edge frequencies.
– Checkbox is not selected:
The list of frequencies yields a set of logarithmically equidistant
frequencies.
N O T E
Alternatively, you can set the Number of Steps on the Standard tab.
In
Manual
mode, you can edit the automatically calculated frequencies.
You can also enter a single frequency or your own list of frequencies.
For example, you can jump between low and high jitter frequencies.
Enter the frequencies in Hz, separated by commas, for example:
2e4, 1.75e6, 6.334e5, ...
7
Jitter Tolerance Tests
370
Agilent J-BERT N4903 High-Performance Serial BERT
Auto mode
Manual mode
Summary of Contents for J-BERT N4903
Page 1: ...S Agilent J BERT N4903 High Performance Serial BERT User Guide s Agilent Technologies...
Page 68: ...2 Setting up Patterns 68 Agilent J BERT N4903 High Performance Serial BERT...
Page 158: ...4 Setting up the Error Detector 158 Agilent J BERT N4903 High Performance Serial BERT...
Page 314: ...6 Evaluating Results 314 Agilent J BERT N4903 High Performance Serial BERT...
Page 374: ...7 Jitter Tolerance Tests 374 Agilent J BERT N4903 High Performance Serial BERT...
Page 394: ...8 Solving Problems 394 Agilent J BERT N4903 High Performance Serial BERT...
Page 434: ...Index 434 Agilent J BERT N4903 High Performance Serial BERT...