Fast Eye Mask - Reference
The Fast Eye Mask measurement returns the results in a numerical
form. The following sections provide explanations of the measured
parameters and the display options that are specific to this
measurement.
Properties that can be specified on the various tabs of the
Properties
dialog box:
Parameters Tab
Note that if you modify the parameters on this page, you have to rerun
the measurement to update the results.
Set the criteria for moving to the next sample point:
•
Number of Compared Bits
After this number of compared bits, the measurement stops for the
current sample point and moves to the next one.
The default is 1 million bits. That means, you can measure a bit error
rate down to 10
-6
(one error per million).
A smaller number reduces the duration of the whole Fast Eye Mask
measurement. A larger number increases the precision of the
measured bit error rates.
•
Number of Errors
After this number of errors, the measurement stops for the current
sample point and moves to the next one. Thisallows you to speed up
the measurement. You can switch off this option if only the number
of compared bits is important.
N O T E
The measurement moves to the next sample point when the first of the two criteria
is reached.
5
Advanced Analysis
246
Agilent J-BERT N4903 High-Performance Serial BERT
Summary of Contents for J-BERT N4903
Page 1: ...S Agilent J BERT N4903 High Performance Serial BERT User Guide s Agilent Technologies...
Page 68: ...2 Setting up Patterns 68 Agilent J BERT N4903 High Performance Serial BERT...
Page 158: ...4 Setting up the Error Detector 158 Agilent J BERT N4903 High Performance Serial BERT...
Page 314: ...6 Evaluating Results 314 Agilent J BERT N4903 High Performance Serial BERT...
Page 374: ...7 Jitter Tolerance Tests 374 Agilent J BERT N4903 High Performance Serial BERT...
Page 394: ...8 Solving Problems 394 Agilent J BERT N4903 High Performance Serial BERT...
Page 434: ...Index 434 Agilent J BERT N4903 High Performance Serial BERT...