Activate this option to configure the error detector to accumulate
over repeated accumulation periods, whereby one period follows
another immediately. There is no "dead time" between the end of one
period and the start of the next.
N O T E
Repeat will not create multiple measurement log files. The data for all repeating
periods will be saved to the same log file.
The accumulation mode cannot be changed while accumulation is running.
Measurement Log
•
Prompt for File Name
Click this option if you want to be prompted for a measurement log
file name before each accumulation period begins. With this option
you can select a new file name for each test run.
•
File Name
Click this option if you want to enter a file name for measurement
logs before starting the accumulation. With this option you will
overwrite the measurement log file with each new test run. However,
in
Repeat
mode, the results of all repetitions will be appended to this
file and no information is lost.
•
No Logging
Click this option to turn off measurement logging. Serial BERT will
not save the test results.
Period
The accumulation period cannot be specified for
Manual
tests.
•
Time
Select this option to configure the error detector to accumulate bit
errors for a specific period of time.
You can enter the desired time period in the
Days
,
Hr
(Hours),
Mn
(Minutes), and
Sec
(Seconds) fields. You may use the numeric keypad
or front panel knob. The minimum value is 1 second and the
maximum value is 99 days, 23 hours, 59 minutes, and 59 seconds.
When the selected time has elapsed, the accumulation ends.
•
Number of Errors
Select this option to configure the error detector to accumulate bit
errors until a specified number of errors has been measured.
Setting up the Error Detector
4
Agilent J-BERT N4903 High-Performance Serial BERT
153
Summary of Contents for J-BERT N4903
Page 1: ...S Agilent J BERT N4903 High Performance Serial BERT User Guide s Agilent Technologies...
Page 68: ...2 Setting up Patterns 68 Agilent J BERT N4903 High Performance Serial BERT...
Page 158: ...4 Setting up the Error Detector 158 Agilent J BERT N4903 High Performance Serial BERT...
Page 314: ...6 Evaluating Results 314 Agilent J BERT N4903 High Performance Serial BERT...
Page 374: ...7 Jitter Tolerance Tests 374 Agilent J BERT N4903 High Performance Serial BERT...
Page 394: ...8 Solving Problems 394 Agilent J BERT N4903 High Performance Serial BERT...