Agilent InfinityLab ELSD User Manual
73
6
Troubleshooting and Diagnostics
General Problems
Cooled Evaporator will not reach low temperature
e.g. 10 °C
High backpressure from detector
Probable cause
Suggested actions
1
Ambient lab temperature too high
Move detector to laboratory where ambient
temperature is <25 °C.
2
Faulty Peltier cooler unit
Consult Agilent Technologies or your local agent
if problem persists.
Probable cause
Suggested actions
1
Nebulizer or nebulizer inlet tube blocked
Manually syringe water into ELSD front Inlet port
to remove obstruction.