Agilent InfinityLab ELSD User Manual
51
4
Using the Module
Operational Parameters
Data Output Rate (Hz)
The rate at which the ELS Detector outputs data can be selected from the
submenu screen. A 10 Hz output rate is selectable for standard LC applications,
a 40 Hz output rate can be chosen for faster LC separations, and 80 Hz is used
for UHPLC type applications with very narrow peaks. The data rate is stored in
memory and is retained even after a power on/off cycle. The default value is
40 Hz.
LAN (TCP/IP Settings)
This screen displays the ELSD's current configured TCP/IP settings. See chapter
LAN Configuration
for details on how to set these values.
With serial connection screen option displays “Instrument Configured for
RS-232”.
Figure 10
TCP/IP settings
NOTE
All ELSD conditions are retained on power cycling of the ELSD detector.