98
Agilent 1290 Infinity Binary Pump User Manual
7
Error Information
Pump Error Messages
Pump drive stroke blocked
Error ID: 22441
The pump drive movement is blocked.
•
Parameter: 1 – 4 referring to pump drive
Pump drive stop not found
Error ID: 22442
The pump drive stop has not been found.
•
Parameter: 1 – 4 referring to pump drive
Pressure sensor calibration wrong or missing
Error ID: 22443
Pressure sensor calibration wrong or missing.
•
Parameter: None
Probable cause
Suggested actions
1
Pump drive motor is mechanically blocked.
Please contact your Agilent service
representative.
Probable cause
Suggested actions
1
Pump drive spindle is defect.
Please contact your Agilent service
representative.
Probable cause
Suggested actions
1
Pressure sensor calibration wrong or
missing.
•
Replace pressure sensor.
•
Please contact your Agilent service
representative.