Agilent 1290 Infinity Binary Pump User Manual
7
Error Information
Pump Error Messages
Waste counter limit exceeded
Error ID: 22056
The counter for the waste volume has exceeded the limit, which has been set
in the user interface.
•
Parameter: None
Flow rate limit exceeded
Error ID: 22064
The flow rate of the binary pump has exceeded the limit, while the pump runs
in pressure controlled mode, e.g. during a pressure test.
•
Parameter: None
Probable cause
Suggested actions
1
The waste container is full.
Empty waste container.
2
Inappropriate setting for waste counter.
•
Reset waste counter.
•
Adjust waste counter limit.
Probable cause
Suggested actions
1
Leak
Check for leaks in the pump and flow path.
2
Bottle empty.
Fill solvent bottle.
3
Shutoff valve closed.
Open shutoff valve.
4
Drift of pressure sensor (unlikely for short
tests taking some minutes).
Replace pressure sensor.