Agilent 1290 Infinity Binary Pump User Manual
83
Error Information
7
General Error Messages
Leak
Error ID: 0064
A leak was detected in the module.
The signals from the two temperature sensors (leak sensor and
board-mounted temperature-compensation sensor) are used by the leak
algorithm to determine whether a leak is present. When a leak occurs, the leak
sensor is cooled by the solvent. This changes the resistance of the leak sensor
which is sensed by the leak-sensor circuit on the main board.
Open Cover
Error ID: 0205
The top foam has been removed.
Probable cause
Suggested actions
1
Loose fittings.
Ensure all fittings are tight.
2
Broken capillary.
Exchange defective capillaries.
Probable cause
Suggested actions
1
Foam not activating the sensor.
Please contact your Agilent service
representative.
2
Dirty or defective sensor.
Please contact your Agilent service
representative.