![Agilent Technologies Infinity Binary Pump VL 1290 User Manual Download Page 204](http://html1.mh-extra.com/html/agilent-technologies/infinity-binary-pump-vl-1290/infinity-binary-pump-vl-1290_user-manual_2867763204.webp)
204
Agilent 1290 Infinity Binary Pump User Manual
12
Hardware Information
Firmware Description
Firmware Description
The firmware of the instrument consists of two independent sections:
•
a non-instrument specific section, called
resident system
•
an instrument specific section, called
main system
Resident System
This resident section of the firmware is identical for all Agilent
1100/1200/1220/1260/1290 series modules. Its properties are:
•
the complete communication capabilities (CAN, LAN and RS-232C)
•
memory management
•
ability to update the firmware of the 'main system'
Main System
Its properties are:
•
the complete communication capabilities (CAN, LAN and RS-232C)
•
memory management
•
ability to update the firmware of the 'resident system'
In addition the main system comprises the instrument functions that are
divided into common functions like
•
run synchronization through APG remote,
•
error handling,
•
diagnostic functions,
•
or module specific functions like
•
internal events such as lamp control, filter movements,
•
raw data collection and conversion to absorbance.