6000 Series Oscilloscope User’s Guide
6
Displaying Data
To decode FlexRay data
To decode FlexRay data, you need a four-channel 6000 Series
mixed signal oscilloscope, the N5432A FlexRay triggering and
decode option, the DECOMSYS::BUSDOCTOR 2 protocol
analyzer, and an MSO communications cable (available from
DECOMSYS).
1
Set up the oscilloscope and BUSDOCTOR 2. See
the Oscilloscope and BUSDOCTOR 2”
2
Set up the trigger mode as described in
You can use FlexRay trigger or another trigger type, such as
duration, I
2
C, Nth edge burst, sequence, SPI, or USB.
For FlexRay triggering set up, see
3
Select the FlexRay serial decode mode:
a
Press the
Acquire
key.
b
Press the
Serial Decode
softkey.
c
In the Serial Decode Menu, press the
Mode
softkey.
d
Use the Entry knob (or repeatedly press and release the
Mode
softkey) to select the
FlexRay
serial decode mode.
Display
decode
Decode
mode
BUSDOCTOR
menu
Return to
previous menu
Reset
counters
N O T E
When the oscilloscope is in FlexRay trigger mode, only FlexRay decode is
available.
Summary of Contents for InfiniiVision DSO6014L
Page 2: ...Agilent Technologies Agilent 6000 Series Oscilloscopes User s Guide...
Page 59: ...58 6000 Series Oscilloscope User s Guide 1 Getting Started...
Page 109: ...108 6000 Series Oscilloscope User s Guide 2 Front Panel Controls...
Page 125: ...124 6000 Series Oscilloscope User s Guide 3 Viewing and Measuring Digital Signals...
Page 251: ...250 6000 Series Oscilloscope User s Guide 5 Making Measurements...
Page 287: ...286 6000 Series Oscilloscope User s Guide 6 Displaying Data Interpreting Decoded LIN Data...
Page 321: ...320 6000 Series Oscilloscope User s Guide 7 Saving and Printing Data...
Page 370: ......