Displaying Data
6
6000 Series Oscilloscope User’s Guide
277
bus is measured. This controls the point within the bit’s time
where the bit value is captured.
8
Set the trigger level.
•
For analog channels, turn the Trigger Level knob.
•
For digital channels, press the
D15 Thru D0
key and the
Thresholds
softkey to access the threshold level setting
softkeys.
9
Return to the previous menu by pressing the up-arrow
softkey.
10
If the decode line does not appear on the display, press the
Decode
softkey to turn it on.
11
If the oscilloscope is stopped, press the
Run/Stop
key to
acquire and decode data.
You can use the
Delayed
horizontal sweep mode for easier
navigation of the decoded data.
One Bit
60%
70%
80%
Sample Point
N O T E
Changing the Source settings in the Serial Decode setup also changes
them in the Trigger setup.
N O T E
If the setup does not produce a stable trigger, the CAN signal may be slow
enough that the oscilloscope is AutoTriggering. Press the Mode/Coupling
key, then press the Mode softkey to set the trigger mode from Auto to
Normal.
Summary of Contents for InfiniiVision DSO6014L
Page 2: ...Agilent Technologies Agilent 6000 Series Oscilloscopes User s Guide...
Page 59: ...58 6000 Series Oscilloscope User s Guide 1 Getting Started...
Page 109: ...108 6000 Series Oscilloscope User s Guide 2 Front Panel Controls...
Page 125: ...124 6000 Series Oscilloscope User s Guide 3 Viewing and Measuring Digital Signals...
Page 251: ...250 6000 Series Oscilloscope User s Guide 5 Making Measurements...
Page 287: ...286 6000 Series Oscilloscope User s Guide 6 Displaying Data Interpreting Decoded LIN Data...
Page 321: ...320 6000 Series Oscilloscope User s Guide 7 Saving and Printing Data...
Page 370: ......