Front-Panel Controls
2
6000 Series Oscilloscope User’s Guide
105
probes do not need to be calibrated, but InfiniiMax probes can
be calibrated. For more information about calibrating probes
see
Results:
User Cal date:
Change in temperature since last User Cal:
Failure:
Comments:
Probe Cal Status:
Self Test
Pressing
Utility
&
Service
&
Start Self Test
performs a series of
internal procedures to verify that the oscilloscope is operating
properly.
It is recommended you run Self Test:
•
after experiencing abnormal operation.
•
for additional information to better describe an oscilloscope
failure.
•
to verify proper operation after the oscilloscope has been
repaired.
Successfully passing Self Test does not guarantee 100% of the
oscilloscope's functionality. Self Test is designed to provide an
80% confidence level that the oscilloscope is operating properly.
About Oscilloscope
Pressing
Utility
&
Service
&
About Oscilloscope
displays information
about your oscilloscope model number, serial number, software
version, boot version, graphics version, and installed licenses.
Summary of Contents for InfiniiVision DSO6014L
Page 2: ...Agilent Technologies Agilent 6000 Series Oscilloscopes User s Guide...
Page 59: ...58 6000 Series Oscilloscope User s Guide 1 Getting Started...
Page 109: ...108 6000 Series Oscilloscope User s Guide 2 Front Panel Controls...
Page 125: ...124 6000 Series Oscilloscope User s Guide 3 Viewing and Measuring Digital Signals...
Page 251: ...250 6000 Series Oscilloscope User s Guide 5 Making Measurements...
Page 287: ...286 6000 Series Oscilloscope User s Guide 6 Displaying Data Interpreting Decoded LIN Data...
Page 321: ...320 6000 Series Oscilloscope User s Guide 7 Saving and Printing Data...
Page 370: ......