Commands by Subsystem
5
Agilent InfiniiVision 7000A Series Oscilloscopes Programmer's Guide
511
:TRIGger:FLEXray:FRAMe:CCBase
Command Syntax
:TRIGger:FLEXray:FRAMe:CCBase <cycle_count_base>
<cycle_count_base> ::= integer from 0-63
The :TRIGger:FLEXray:FRAMe:CCBase command sets the base of the
FlexRay cycle count (in the frame header) to trigger on. The cycle count
base setting is only valid when the FlexRay trigger mode is set to FRAME.
Query Syntax
:TRIGger:FLEXray:FRAMe:CCBase?
The :TRIGger:FLEXray:FRAMe:CCBase? query returns the current cycle
count base setting for the FlexRay frame trigger setup.
Return Format
<cycle_count_base><NL>
<cycle_count_base> ::= integer from 0-63
See Also
•
"Introduction to :TRIGger Commands"
•
N O T E
This command is only valid when the FLEXray triggering and serial decode option (Option
FLX) has been licensed.