:TRIGger[:EDGE]:SLOPe
Command Syntax
:TRIGger[:EDGE]:SLOPe <slope>
<slope> ::= {NEGative | POSitive | EITHer | ALTernate}
The :TRIGger[:EDGE]:SLOPe command specifies the slope of the edge for
the trigger. The SLOPe command is not valid in TV trigger mode. Instead,
use :TRIGger:TV:POLarity to set the polarity in TV trigger mode.
Query Syntax
:TRIGger[:EDGE]:SLOPe?
The :TRIGger[:EDGE]:SLOPe? query returns the current trigger slope.
Return Format
<slope><NL>
<slope> ::= {NEG | POS | EITH | ALT}
See Also
•
"Introduction to :TRIGger Commands"
•
•
Example Code
' TRIGGER_EDGE_SLOPE - Sets the slope of the edge for the trigger.
' Set the slope to positive.
myScope.WriteString ":TRIGGER:EDGE:SLOPE POSITIVE"
Example program from the start:
Summary of Contents for InfiniiVision 7000A Series
Page 1: ...Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide...
Page 34: ...34 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 1 What s New...
Page 44: ...44 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 2 Setting Up...
Page 58: ...58 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 3 Getting Started...
Page 750: ...750 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 8 Error Messages...
Page 922: ...922 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide Index...