Agilent InfiniiVision 7000A Series Oscilloscopes Programmer's Guide
5
Commands by Subsystem
:MEASure:PWIDth
Command Syntax
:MEASure:PWIDth [<source>]
<source> ::= {<digital channels> | CHANnel<n> | FUNCtion | MATH}
<digital channels> ::= DIGital0,..,DIGital15 for the MSO models
<n> ::= {1 | 2 | 3 | 4} for the four channel oscilloscope models
<n> ::= {1 | 2} for the two channel oscilloscope models
The :MEASure:PWIDth command installs a screen measurement and starts
the positive pulse width measurement. If the optional source parameter is
specified, the current source is modified.
Query Syntax
:MEASure:PWIDth? [<source>]
The :MEASure:PWIDth? query measures and outputs the width of the
displayed positive pulse closest to the trigger reference. Pulse width is
measured at the midpoint of the upper and lower thresholds.
IF the edge on the screen closest to the trigger is falling:
THEN width = (time at trailing falling edge - time at leading rising edge)
ELSE width = (time at leading falling edge - time at leading rising edge)
Return Format
<value><NL>
<value> ::= width of positive pulse in seconds in NR3 format
See Also
•
"Introduction to :MEASure Commands"
•
•
•
N O T E
This command is not available if the source is FFT (Fast Fourier Transform).
Summary of Contents for InfiniiVision 7000A Series
Page 1: ...Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide...
Page 34: ...34 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 1 What s New...
Page 44: ...44 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 2 Setting Up...
Page 58: ...58 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 3 Getting Started...
Page 750: ...750 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 8 Error Messages...
Page 922: ...922 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide Index...