Commands by Subsystem
5
Agilent InfiniiVision 7000A Series Oscilloscopes Programmer's Guide
171
:OVLRegister (Overload Event Register)
Query Syntax
:OVLRegister?
The :OVLRegister query returns the overload protection value stored in the
Overload Event Register (OVLR). If an overvoltage is sensed on a 50
Ω
input, the input will automatically switch to 1 M
Ω
input impedance. A "1"
indicates an overload has occurred.
N O T E
You can set analog channel input impedance to 50
Ω
on the 300 MHz, 500 MHz, and 1 GHz
bandwidth oscilloscope models. On these same bandwidth models, if there are only two
analog channels, you can also set external trigger input impedance to 50
Ω
.
9
8
10
11
12
13
14
15
To OVLR bit in
O pe ra ti o n Status
Register
+
OR
1
0
2
3
4
5
6
7
:OVLR?
Overload Event Register
:OVL
:OVL?
Overload Event Enable
(Mask) Register
Chan2
OVL
Chan1
OVL
Chan3
OVL
Chan4
OVL
Ext Trig
OVL
Chan2
Fault
Chan1
Fault
Chan3
Fault
Chan4
Fault
Ext Trig
Fault
Table 55 Overload Event Register (OVLR)
Bit
Description
When Set (1 = High = True), Indicates:
15-11
---
(Not used.)
10
External Trigger Fault
Fault has occurred on External Trigger input.
9
Channel 4 Fault
Fault has occurred on Channel 4 input.
8
Channel 3 Fault
Fault has occurred on Channel 3 input.
7
Channel 2 Fault
Fault has occurred on Channel 2 input.
6
Channel 1 Fault
Fault has occurred on Channel 1 input.
5
---
(Not used.)
Summary of Contents for InfiniiVision 7000A Series
Page 1: ...Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide...
Page 34: ...34 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 1 What s New...
Page 44: ...44 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 2 Setting Up...
Page 58: ...58 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 3 Getting Started...
Page 750: ...750 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 8 Error Messages...
Page 922: ...922 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide Index...