120
Agilent InfiniiVision 7000A Series Oscilloscopes Programmer's Guide
5
Commands by Subsystem
*ESR (Standard Event Status Register)
Query Syntax
*ESR?
The *ESR? query returns the contents of the Standard Event Status
Register. When you read the Event Status Register, the value returned is
the total bit weights of all of the bits that are high at the time you read
the byte. Reading the register clears the Event Status Register.
The following table shows bit weight, name, and condition for each bit.
Table 42 Standard Event Status Register (ESR)
Bit
Name
Description
When Set (1 = High = True), Indicates:
7
PON
Power On
An OFF to ON transition has occurred.
6
URQ
User Request
A front-panel key has been pressed.
5
CME
Command Error
A command error has been detected.
4
EXE
Execution Error
An execution error has been detected.
3
DDE
Device Dependent Error
A device-dependent error has been detected.
2
QYE
Query Error
A query error has been detected.
To ESB bit in
Status Byte
Register
Standard Event
Status Enable
(Mask) Register
Standard Event
Status Register
*ESR?
*ESE
*ESE?
PON
URQ
EXE
DDE
QYE
RQL
OPC
CME
+
OR
1
0
2
3
4
5
6
7
Summary of Contents for InfiniiVision 7000A Series
Page 1: ...Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide...
Page 34: ...34 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 1 What s New...
Page 44: ...44 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 2 Setting Up...
Page 58: ...58 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 3 Getting Started...
Page 750: ...750 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 8 Error Messages...
Page 922: ...922 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide Index...