:SBUS<n>:I2S:TRIGger:PATTern:DATA
Command Syntax
:SBUS<n>:I2S:TRIGger:PATTern:DATA <string>
<string> ::= "n" where n ::= 32-bit integer in signed decimal when
<base> = DECimal
<string> ::= "nn...n" where n ::= {0 | 1 | X | $} when
<base> = BINary
<string> ::= "0xnn...n" where n ::= {0,..,9 | A,..,F | X | $} when
<base> = HEX
The :SBUS<n>:I2S:TRIGger:PATTern:DATA command specifies the I2S
trigger data pattern searched for in each I2S message.
Set a <string> bit to "0" or "1" to set the corresponding bit in the data
pattern to low or high, respectively.
Set a <string> bit to "X" to ignore (mask off) that bit in the data pattern.
Use the "$" character to indicate that the value of the corresponding bit
will not be changed (the existing bit value is used).
When <base> = DECimal, the "X" and "$" characters cannot be entered.
When queried, the "$" character is returned when any bits in the pattern
have the value of "X" and <base> = DECimal. When any bits in a given
nibble have the value of "X" and <base> = HEX, the "$" character is
returned for the corresponding nibble.
N O T E
<base> is specified with the :SBUS<n>:I2S:TRIGger:PATTern:FORMat command. The
default <base> is DECimal.
N O T E
The :SBUS<n>:I2S:TRIGger:PATTern:DATA command specifies the I2S trigger data pattern
used by the EQUal, NOTequal, GREaterthan, and LESSthan trigger conditions. If the
GREaterthan or LESSthan trigger condition is selected, the bits specified to be masked off
("X") will be interpreted as 0's.
N O T E
The length of the trigger data value is determined by the :SBUS<n>:I2S:RWIDth and
:SBUS<n>:I2S:TWIDth commands. When the receiver word size is less than the transmitter
word size, the data length is equal to the receiver word size. When the receiver word size is
greater than the transmitter word size, the data length is equal to the transmitter word size.
Summary of Contents for InfiniiVision 3000 X-Series
Page 1: ...s1 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide...
Page 30: ...30 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide...
Page 54: ...54 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 2 Setting Up...
Page 68: ...68 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 3 Getting Started...
Page 218: ...218 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 6 Root Commands...
Page 242: ...242 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 8 BUS n Commands...
Page 280: ...280 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 11 DEMO Commands...
Page 288: ...288 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 12 DIGital d Commands...
Page 340: ...340 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 15 FUNCtion Commands...
Page 358: ...358 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 16 HARDcopy Commands...
Page 378: ...378 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 18 MARKer Commands...
Page 446: ...446 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 19 MEASure Commands...
Page 564: ...564 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 23 POWer Commands...
Page 842: ...842 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 29 TIMebase Commands...
Page 1054: ...1054 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 35 Error Messages...
Page 1076: ...1076 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 36 Status Reporting...
Page 1232: ...1232 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide Index...