:SBUS<n>:CAN:TRIGger
Command Syntax
:SBUS<n>:CAN:TRIGger <condition>
<condition> ::= {SOF | DATA | ERRor | IDData | IDEither | IDRemote |
ALLerrors | OVERload | ACKerror}
The :SBUS<n>:CAN:TRIGger command sets the CAN trigger on condition:
•
SOF - will trigger on the Start of Frame (SOF) bit of a Data frame,
Remote Transfer Request (RTR) frame, or an Overload frame.
•
DATA - will trigger on CAN Data frames matching the specified Id,
Data, and the DLC (Data length code).
•
ERRor - will trigger on CAN Error frame.
•
IDData - will trigger on CAN frames matching the specified Id of a
Data frame.
•
IDEither - will trigger on the specified Id, regardless if it is a Remote
frame or a Data frame.
•
IDRemote - will trigger on CAN frames matching the specified Id of a
Remote frame.
•
ALLerrors - will trigger on CAN active error frames and unknown bus
conditions.
•
OVERload - will trigger on CAN overload frames.
•
ACKerror - will trigger on a data or remote frame acknowledge bit that
is recessive.
The table below shows the programming parameter and the corresponding
front- panel softkey selection:
Remote <condition> parameter
Front-panel Trigger on: softkey selection
(softkey text - softkey popup text)
SOF
SOF - Start of Frame
DATA
ID & Data - Data Frame ID and Data
ERRor
Error - Error frame
IDData
ID & ~RTR - Data Frame ID (~RTR)
IDEither
ID - Remote or Data Frame ID
IDRemote
ID & RTR - Remote Frame ID (RTR)
ALLerrors
All Errors - All Errors
OVERload
Overload - Overload Frame
ACKerror
Ack Error - Acknowledge Error
Summary of Contents for InfiniiVision 3000 X-Series
Page 1: ...s1 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide...
Page 30: ...30 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide...
Page 54: ...54 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 2 Setting Up...
Page 68: ...68 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 3 Getting Started...
Page 218: ...218 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 6 Root Commands...
Page 242: ...242 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 8 BUS n Commands...
Page 280: ...280 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 11 DEMO Commands...
Page 288: ...288 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 12 DIGital d Commands...
Page 340: ...340 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 15 FUNCtion Commands...
Page 358: ...358 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 16 HARDcopy Commands...
Page 378: ...378 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 18 MARKer Commands...
Page 446: ...446 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 19 MEASure Commands...
Page 564: ...564 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 23 POWer Commands...
Page 842: ...842 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 29 TIMebase Commands...
Page 1054: ...1054 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 35 Error Messages...
Page 1076: ...1076 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 36 Status Reporting...
Page 1232: ...1232 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide Index...