4-22
N9360A Multi UE Tester GSM Programming Manual
4
Programming Command Guideline
Standard Event Status Enable Register
The Standard Event Status Enable Register is used to mask the
Standard Event Status Register information to inform summary
information to the Status Byte Register.
Figure 4-9 Standard Event Status Enable Register
To inform the summary information of the Standard Event
Status Register to bit 5 of the Status Byte Register, set
1
to
corresponding bit of the Standard Event Status Enable Register.
Send the
*ESE <int>
command, where
<int>
is the sum of
the decimal values of the bits you want to enable. For example,
to inform the Power On and Command Error bits information to
the bit 5 of the Status Byte Register, bit 7 and 5 are set to
160
by
*ESE 160
command (128 plus 32). See
. The query
command
*ESE?
returns the decimal value of the sum of the
bits previously set by with the
*ESE <int>
command.