Programming Command Reference
5
N9360A Multi UE Tester GSM Programming Manual
5-157
BER Measurement
GSM mode, Test Loop C
TESTs:MANual:RESults:SENSitivity:BERFast?
GPRS mode, Test Mode B
TESTs:MANual:RESults:SENSitivity:BERGprs?
EGPRS mode, Test Mode SRB
TESTs:MANual:RESults:SENSitivity:SRB?
This query command returns the Sensitivity measurement
result values
<int 1>
,
<int 2>
and
<real 3>
. Each value is
separated by a comma.
•
<int 1>
(1 digit) is
0
if measurement results are within the
measurement range, or
1
if any measurement result is out of
range or the BER is not measured.
•
<int 2>
(1digit) is the test result for the
0
(pass) or
1
(fail).
•
<real 3>
is the measured result of the BER data ranging
from
0.00
to
100.00
in %. If measurement result is out of
range or the BER is not measured, the value
99.99
is
returned.
Sensitivity Test Frames
GSM Mode, Test Loop A
TESTs:MANual:SENSitivity:FRAMe <int>
TESTs:MANual:SENSitivity:FRAMe?
GSM Mode, Test Loop C
TESTs:MANual:SENSitivity:FRAMe:BERFast <int>
TESTs:MANual:SENSitivity:FRAMe:BERFast?
N O T E
The above commands are valid only for corresponding mode setting.