Agilent InfinityLab LC Series Vialsamplers User Manual
94
3
Using the Module
Install the Optional Integrated Column Compartment
Connect a Column Identification Tag to the Tag
Reader
Tools required
p/n
Description
8710-0510
Open-end wrench 1/4 — 5/16 inch
Parts required
p/n
Description
5067-5917
InfinityLab Column Identification Tag
Column (up to 30 cm length)
Preparations
•
Install the Column ID Tag Reader (see
“Install the Column ID Upgrade Kit”
•
Remove the column if installed (see
“Remove a Column from the ICC”
1
If the column is without a preinstalled identification tag,
fit a new tag onto it by slipping the lanyard loop over one
end of the column and pulling the cord tight through the
plastic holder.
NOTE
Once the lanyard is fixed, the tag can no longer be removed
from the column.
2
Install the column in the ICC (see